HILLROM PROGRESSA+ BED, Product Codes: P7501A000016, P7501A000019, P7501A000031, P7501A000032, P7501A000033, P7501A000052, P7501A000053, P7501A000054, P7501A000055, P7501A001851, P7501A001972, P7501A001973, and P7501A001974

FDA Recall Z-0212-2024 · Posted 2023-11-02

Recall StatusOpen, Classified
Product CodeIOQ
Date Initiated2023-10-04
Product Quantity499 beds

Reason for Recall

Progressa+ beds have the potential for static buildup from the casters and/or power drive. This static electricity could be passed to the operator through the transport handle during transport, resulting in a mild static electric shock.

Firm Action

Baxter issued an Urgent Medical Device Correction notice to its consignees on 10/04/2023 by USPS first class mail. The notice explained the potential problem and hazard. Baxter will be correcting this issue by installing a conductive drag chain on the Progressa+ bed frame to prevent the electrostatic charge buildup. A Baxter representative will contact your facility to schedule a device correction to add the conductive drag chain. Please note you will be receiving this correction from Baxter at no charge. All units purchased in the future will have the drag chain.

Root Cause

Under Investigation by firm

Distribution Pattern

US Nationwide Distribution in the states of: CA, FL, IN, MA, NY, OH, TX, VA