LEAD 3387S-40 STIMLOC DBS, Model Number 3387S-40. Electrical implantable device.
FDA Recall Z-0217-2024 · Posted 2023-11-03
Advertisement
Recalling FirmMedtronic Neuromodulation
Recall StatusCompleted
Product CodeMHY
Date Initiated2022-05-27
Product Quantity14 units
Reason for Recall
There are a small number of units of the DBS Lead with StimLoc kit (3387S-40) due to a discrepancy in the Use By Date (UBD) printed on the outer packaged kit label versus the UBD printed on the individual product label.
Firm Action
Medtronic issued a Medical Device Removal notice to its consignees on 05/27/2023 by US mail. The notice explained the issue, potential risk, and requested the return of the affected product.
Root Cause
Labeling Change Control
Distribution Pattern
US: MO, FL, TX, WA