Terminated

TOSOH brand AIA-600 II Enzyme Immunoassay Analyzer; Software version 3.12, Catalog Numbers 019359, 019400; Model 019400, Smart Media

FDA Recall Z-0218-05 · Posted 2004-11-30

Recall StatusTerminated
Product CodeNIG
Date Initiated2003-04-02
Date Terminated2005-06-15
Product Quantity81 units

Reason for Recall

A software upgrade was released by the firm for its AIA-600 II analyzers, in that washing deficiencies caused by the analyzer were corrected.

Firm Action

The firm initiated the correction on 4/2/03 via telephone calls and PR on 11/24/04 to all its direct customers, however, the firm indicated that software upgrade was accomplished by the firm's Field Service Engineers.

Root Cause

Other

Distribution Pattern

81 customers received the upgrade, including customers who received instruments with the upgrade installed already. 37 customers received the v.3.12 installation. In addition, TOSOH bioscience has identified 3 foreign and 5 US customers who required the upgrade. The firm had distributed to 3 US Government medical centers as well. These upgrades have been completed.