Terminated

10 mm/135 degree Titanium Cannulated Trochanteric Fixation Nail 170 MM - Sterile: intended to treat stable and unstable fractures of the proximal femur

FDA Recall Z-0219-2015 · Posted 2014-11-14

Recalling FirmSynthes, Inc.
Recall StatusTerminated
Product CodeHSB
Date Initiated2014-10-29
Date Terminated2015-12-18
Product Quantity6

Reason for Recall

The 10 mm/135 degree Titanium Cannulated Trochanteric Fixation Nail 170 MM (Sterile) was assembled incorrectly. If the non conformance goes undetected, it may impact the locking function post-operatively and compromise the bone reduction and construct stability which may lead to non-union/malunion.

Firm Action

An urgent medical device recall notice, dated October 29, 2014, was sent to end users that identified the product, problem and action to be taken. Customers were instructed to return the device to Synthes along with completing the verification section of the notice.

Root Cause

Employee error

Distribution Pattern

US Distribution in the states of: AK, FL, OH, OK, and TX.