AD-TECH Spencer Probe Depth Electrode, REF SD06R-AP58X-000 6-Contact 1.12mm (A style) Depth Electrode
FDA Recall Z-0220-2025 · Posted 2024-10-24
Advertisement
Recalling FirmAd-Tech Medical Instrument Corporation
Recall StatusCompleted
Product CodeGZL
Date Initiated2024-09-19
Product Quantity4 units
Reason for Recall
Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.
Firm Action
AD-TECH issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 09/19/2024. The notice explained the problem with the devices, risk to health, and requested the consignee contact customer support to obtain an RMA to expedite the return of the affected product.
For questions: contact your Ad-Tech Clinical Specialist.
Root Cause
Employee error
Distribution Pattern
US Nationwide distribution in the states of CA, OH.
Related FDA 510(k) Records