Wagner cone Prosthesis, uncemented, sterile, Rx
FDA Recall Z-0221-2013 · Posted 2012-11-06
Advertisement
Recalling FirmZimmer, Inc.
Recall StatusTerminated
Product CodeLZO
Date Initiated2012-09-04
Date Terminated2014-04-16
Product Quantity56
Reason for Recall
Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 20I0 by the Zimmer GmbH production site in Winterthur, Switzerland, after receiving four complaints of a discrepancy between product patient labels and product labels with respect to the reference and lot numbers of the product. While the information printed on the product label is correct, the info
Firm Action
Zimmer, Inc. sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated September 4, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. A Certificate of Acknowledgement Form was attached for customers to complete and return via fax to Zimmer, Inc. at 574 372-4265.
Root Cause
Error in labeling
Distribution Pattern
Nationwide Distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.