MC-500 Multicolor Laser Photocoagulator using software version 2.20. Opthalmic device used for retinal photocoagulation for treatment of ocular fundus diseases.
FDA Recall Z-0222-2015 · Posted 2014-11-17
Advertisement
Recalling FirmNidek Inc
Recall StatusTerminated
Product CodeHQF
Date Initiated2013-05-24
Date Terminated2014-11-17
Product Quantity5
Reason for Recall
Retrospective review found that a Engineering Change Order initiated in May 2013 for a memory function problem was not reported as required.
Firm Action
Nidek Engineers visited all affected sites to install new software version.
For questions regarding this recall call 510-353-7785.
Root Cause
Software design
Distribution Pattern
US Distribution in the states of: CA, NE, and NY.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.