Nidek Excimer Laser Corneal Surgery System, model EC-5000 CXIII. Opthalmic Device. The system is composed as a laser generator which produces an excimer laser radiation of wavelength 193nm, a beam delivery unit, an optical system for observation, a gas system and a computer for system control.
FDA Recall Z-0223-2015 · Posted 2014-11-17
Advertisement
Recalling FirmNidek Inc
Recall StatusTerminated
Product CodeLZS
Date Initiated2012-08-06
Date Terminated2014-11-17
Product Quantity14 units
Reason for Recall
Contact failure in F1 and F4 fuse holders was detected on the EC-5000 CXIII Excimer Laser Corneal Surgery System that could cause a serious adverse event for the patient.
Firm Action
Nidek Engineers visited each site to perform correction - fuse holders have been inspected and replaced if necessary.
An Engineering Change Order was issued on June 1, 2012, for inspection and/or replacement of affected fuse holders by Nidek Service Engineers.
For further questions please call (510) 353-7785.
Root Cause
Component design/selection
Distribution Pattern
US Distribution including the states of CO, NY, MO, LA, ME, CA, MI, AZ, CT and FL.