Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

NEO-fit Neonatal Endotracheal Tube Grip, Part Number 42-2540

FDA Recall Z-0223-2023 · Posted 2022-11-10

Recalling FirmCooperSurgical, Inc.
Recall StatusOpen, Classified
Product CodeJAY
Date Initiated2022-10-20
Product Quantity220,500 devices

Reason for Recall

Firm has received 11 complaints involving a loose or detached metal clips, including 3 cases where the metal clip was ingested by the patient, and 3 cases where the metal clip was removed from the patient's mouth. In one case, the patient suffered a laceration to the throat.

Firm Action

On October 20, 2022, the firm notified affected customers via "URGENT: MEDICAL DEVICE FIELD SAFETY NOTICE" letters dated 10/7/2022. Customers were instructed to quarantine all affected product and complete the appropriate version of the attached form (Customer Acknowledgement Form or Distributor Acknowledgement Form). Once completed, return the form to CooperSurgical as indicated at the top of the form. Note: Even if you do not have any affected product in your inventory (or have not distributed any affected product to customers), please complete and return the enclosed form so that we may document receipt of this FSN. CooperSurgical will then arrange for the return of the affected product at no additional cost to the customer. Customers will receive a credit for returned product. If you have any questions, please feel free to reach us at 203-601-5200 ext. 3300 or [email protected].

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide distribution: US (Nationwide) including Puerto Rico; and countries (foreign) of: Australia, Bahrain, Belgium, Bulgaria, Canada, Colombia, Cyprus, Ecuador, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italy, Malaysia, Malta, Netherlands, New Zealand, Saudi Arabia, Singapore, Turkey, Philippines, Qatar, United Arab Emirates, and United Kingdom.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑