Roche Cobas driver (rchcob6i), analytical laboratory data interface.
FDA Recall Z-0224-2007 · Posted 2006-11-29
Advertisement
Recalling FirmData Innovations, Inc.
Recall StatusTerminated
Product CodeJQP
Date Initiated2006-11-03
Date Terminated2015-06-11
Product Quantity56
Reason for Recall
Software of modular driver may incorrectly report patient results as Quality Control Results
Firm Action
Data Innovations notified users by email listserv and by letter dated 10/25/06.
Root Cause
Software change control
Distribution Pattern
Nationwide and Canada.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.