Magnetom Trio TIM, magnetic resonance imaging
FDA Recall Z-0226-2007 · Posted 2006-11-29
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeLNH
Date Initiated2006-10-12
Date Terminated2008-04-23
Product Quantity9 units
Reason for Recall
Possible excessive RF exposure/may burn
Firm Action
The recalling firm has issued a Customer Safety Advisory to affected customers per Update Instructions MR038/06/S. The letter informs customers of the potential issue and instructs them not to use the involved coils until this issue can be corrected.
Root Cause
Other
Distribution Pattern
Nationwide.