Synthes Titanium Polyaxial Reduction Head for Titanium Matrix Spine Screws: The Titanium Polyaxial Reduction Head is a part of the Matrix Spine System. Intended for posterior pedicle screw fixation (Tl OS2- ilium), posterior hook fixation (Tl-L5), or anterolateral fixation (T8-L5).

FDA Recall Z-0226-2015 · Posted 2014-11-26

Recall StatusTerminated
Product CodeNKB
Date Initiated2014-10-23
Date Terminated2015-08-10
Product Quantity24

Reason for Recall

Two nonconforming parts of the Titanium Polyaxial Reduction Head for Titanium Matrix Spine Screws may have been inadvertently packaged and shipped. It is believed that the inner diameter of the Reduction Head may not have been machined to the correct specification.

Firm Action

DePuy Synthes sent an Urgent Notice: Medical Device Recall letter, dated October 23, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check inventory and remove affected stock, call Synthes to obtain a Return Authorization for the device, and to return the completed verification section of the response form by fax to (610) 430-7083 or Scan/email: [email protected]. Customers with questions were instructed to call 610-719-5450.

Root Cause

Process control

Distribution Pattern

Distributed to MI, MT, and CO.