VirtuoSaph Plus Endoscopic Vessel Harvesting System, Catalog Number: VSP550EX Product Usage: The VittuoSaph Plus Endoscopic Vessel Harvesting System VSP550EX is indicated for use in minimally invasive surgery allowing access for vessel harvesting, and is primarily indicated for patients undergoing endoscopic surge1y for arterial bypass. It is indicated for cutting tissue and controlling bleeding through coagulation, and for patients requiting blunt dissection of tissue including dissection of blood vessels and dissection of blood vessels of the extremities. Extremity procedures include tissue dissection and/or vessel harvesting along the saphenous vein for coronary artery bypass grafting and peripheral artery bypass grafting or radial artery for use in coronary artery bypass grafting vc

FDA Recall Z-0227-2019 · Posted None

Recall StatusTerminated
Product CodeGEI
Date Initiated2018-08-31
Date Terminated2019-10-21
Product Quantity135

Reason for Recall

The system's IFU contains a list of generators that the have been deemed compatible. Recent reports indicate that one energy platform is not compatible with the system. Therefore, it has been removed from the list of compatible generators within the IFU.

Firm Action

Urgent Medical Device Correction notification letters dated August 29, 2018 were distributed to customers. The letter identified the affected product, problem and actions to be take. Customers were instructed to: Review this Medical Device Correction Notification and the Recommended Actions. Assure that all users receive notice of this issue. Confirm receipt of this correction notification by emailing or faxing the attached Customer Response Form to the email address or fax number indicated on the form. Ensure that users are notified of the revised IFU provided with this notification (Revision 3) and that it replaces any previous versions at all points of use within your facility. If you wish to return Revision 2 of the IFU to Terumo CVS, follow your complaint reporting protocol and contact Terumo CVS. For questions contact Terumo CVS Customer Service 1-800-521-2818.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide - US Nationwide in the states of FL, IL, NY, OH, TX. The products were distributed to the following foreign country: Singapore.