Elekta Disposable Biopsy Needle Kit for Leksell Stereotactic System, REF 911933, Diameter / 14 Gauge, containing 6 inner boxes labeled REF 911761, Disposable Biopsy Needle Kit for Leksell Stereotactic System. Each inner box of REF 911761 contains a Tyvek bag labeled REF 912465, Disposable Biopsy Needle and Aspirator Tube. The system is used for for localization and diagnosis of intracranial disorders and their surgical treatment including radiosurgery and stereotactic radiation therapy.
FDA Recall Z-0227-2025 · Posted 2024-11-06
Advertisement
Recalling FirmElekta Instrument AB
Recall StatusOpen, Classified
Product CodeHAW
Date Initiated2024-09-25
Product Quantity111 kits (666 needles)
Reason for Recall
Elekta has become aware that Disposable Biopsy Needles can contain some microscopic debris on the inside of the biopsy needle. The material in the debris is stainless steel, same material as the biopsy needle. No debris has been found on the outer parts of the biopsy needles.
Firm Action
Elekta issued an URGENT IMPORTANT FIELD SAFETY NOTIFICATION to its consignees on 069/23/2023 via email. Then notice explained the problem, clinical impact, and requested the destruction of the affected product. Contact the local Elekta representative for alternative solutions. They also requested that the notice be posted in a place accessible to all users until this action is closed and to notify all appropriate personnel working with this product of the content of the notice.
Root Cause
Under Investigation by firm
Distribution Pattern
US: AZ, TX, FL, CA, NY, GA, VA, NJ, MO, PA, WA,
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.