Inspiration Ventilator System, i Series and LS 12.1 LCD. The Inspiration 7i / 5i Ventilator System is to be used by healthcare professionals in hospitals or healthcare facilities and intra-hospital transport.

FDA Recall Z-0230-2016 · Posted 2015-11-19

Recalling FirmEvent Medical LTD
Recall StatusTerminated
Product CodeCBK
Date Initiated2015-10-13
Date Terminated2017-06-22
Product Quantity6441 units total (251 units in US)

Reason for Recall

A component on the systems power board can fail in a manner which can cause the ventilator to shut down without alarming.

Firm Action

eVent Medical sent an Urgent Field Safety Notice dated October 12, 2015, to all affected customers on October 13, 2015, to inform them that eVent Medical is recalling Inspiration Ventilators i Series and LS 12.1 LCD because a component on the systems power board can fail in a manner which can cause the ventilator to shut down without alarming. The letter provides the customers with the problems identified and the actions to be taken. Customers with questions are instructed to call (949) 900-1917, x232.

Root Cause

Component design/selection

Distribution Pattern

Worldwide Distribution - US (Nationwide) and Internationally to: Algeria Argentina Australia Austria Bahrain Bangladesh Barbados Belarus Belgium Bolivia Canada Chile China Colombia Czech Rep Ecuador Egypt England Estonia Finland Germany Ghana Gibraltar Greece Guatemala Holland Hong Kong Hungary India Indonesia Iran Iraq Ireland Israel Italy Japan Jordan Kenya Korea Kuwait Laos Latin America Lebanon Libya Lithuania Malaysia Mexico Morocco Nepal Netherlands New Zealand Norway Pakistan Palestine Panama Peru Philippines Poland Portugal Russia Saudi Arabia Serbia Singapore South Africa Spain Sri Lanka Sudan Switzerland Syria Taiwan Thailand Turkey UAE Ukraine USA Venezuela Vietnam Yemen Zimbabwe