Recall StatusOpen, Classified
Date Initiated2022-10-12
Product Quantity175 devices
Reason for Recall
Due to Indirect Decompression System devices exhibiting a potential for weld breakage at the spindle cap-to-main body interface. As a result of the separation of the spindle cap from the main body it may prevent proper deployment of the device during implantation.
Firm Action
On 10/12/2022, the firm, Boston Scientific, sent an "Urgent: Medical Device Recall" letter via FedEx to customers. The letter described the product, problem and actions to be taken. The Customers are instructed to:
-Review their facility's device inventory: Segregate and return any unused Superion IDS devices from the affected subset to Boston Scientific. Complete and return account-specific Reply Verification Tracking Form via Email:
[email protected]
or Fax to: Field Action Center at 1-763-415-7708.
-Patient follow-up: Discuss this advisory with their patient(s) who are implanted with an affected device. Promptly investigate any reported pain and/or discomfort, or any suspected indication of migration of the Superion IDS implant, which may suggest a potential spindle cap weld break. If an affected device remains implanted in their patient, and successful deployment and/or device structural integrity is questioned, consider confirming via imaging.
-If a spindle cap weld break is confirmed post-implant: Assess the clinical benefits versus risks associated with explanting an affected Superion IDS from their patient. Note that the Superion IDS Inserter tool would be unable to engage a non-intact device for explant. Medical judgement is required for determining the appropriate surgical method for explant in such cases.
-Records: For each patient with an affected Superion IDS, append their medical record with a copy of this letter to maintain awareness of this topic for the remaining service life of the device. Should any revision/replacement occur, please inform Boston Scientific with relevant clinical information.
For additional information or questions - contact the local Boston Scientific representative, or the Firm's Public Contact, Director, Global Corporate Communications at 415-730-2310 or
[email protected].
Root Cause
Process change control