Open, Classified

Fuse ULTRA Foot Plating System Instructions for Use. Models: OS200400, OS200401, OS328017, OS200020L, OS200025L, OS200120L, OS200125, OS200125L, OS200135, OS200324, OS200325, OS200326, OS328026, OS328028C, OS216150, OS218150, OS292110, OS292120, OS292160, OS292200, OS200516, OS200518, OS200520, OS200522, OS200616, OS200618, OS200620, OS200622, OS421300, OS421302, OS421304, OS421306, OS421308, OS421208, OS421210, OS421212, OS421000L, OS421000R, OS421002L, OS421002R, OS421004L, OS421004R, OS421006L, OS421006R, OS42260LL, OS42260LR, OS42260ML, OS42260MR, OS42260SL, OS42260SR, OS42260XSL, OS42260XSR, OS422800, OS422803, OS422804, OS422805, OS422807, OS422825, OS422400, OS422403, OS422404, OS422405, OS422406, OS422407, OS421406, OS421408, OS421414, OS422702, OS422703, OS422712, OS422714, OS422716, OS421512, OS421514, OS421516, OS421112, OS421116, OS421120, OS421124, OS421130, OS42150L, OS42150M, OS42150S, OS42150XS, OS326920-T, OS326922-T, OS326924-T, OS326925, OS326925-T, OS326926-T, OS326928, OS326928-T, OS326930, OS326930-T, OS326932, OS326932-T, OS326934, OS326934-T, OS326936, OS326936-T, OS326938, OS326938-T, OS326940, OS326940-T, OS326945, OS326945-T, OS326950, OS326950-T, OS326955, OS326955-T, OS326960, OS326960-T, OS327920, OS327922, OS327924, OS327926, OS327928, OS327930, OS327932, OS327934, OS327936, OS327938, OS327940, OS327945, OS327950, OS327955, OS327960, OS421710, OS421712, OS421714, OS421716, OS421718, OS421720, OS421722, OS421724, OS421726, OS421728, OS421730, OS421735, OS421740, OS421745, OS421750, OS422508, OS422510, OS422512, OS422514, OS422516, OS422518, OS422520, OS422522, OS422524, OS422526, OS422528, OS422530, OS421814, OS421816, OS421818, OS421820, OS421822, OS421824, OS421826, OS421828, OS421830, OS421835, OS421840, OS421845, OS421850, OS422608, OS422610, OS422612, OS422614, OS422616, OS422618, OS422620, OS422622, OS422624, OS422626, OS422628, OS422630, R327924, R327928, R327932, R327936, R327940, R327945, R327950, R421000L, R421000R, R421710, R421712, R421714, R421716, R421718, R421720, R421722, R421724, R421726, R421728, R421730, R421735, R421740, R421814, R421816, R421818, R421820, R421822, R421824, R421826, R421828, R421830, R421835, R421840, R421845, R421850, R42260ML, R42260MR, R42260SL, R42260SR, R42260XSL, R42260XSR

FDA Recall Z-0230-2026 · Posted 2025-10-21

Recall StatusOpen, Classified
Product CodeHRS
Date Initiated2025-09-17
Product Quantity3317

Reason for Recall

Instructions for use for a foot plating system is being updated to remove the second option of requirements for sterilization which was 3 minutes exposure time at 134-137-degrees C (273-279-degrees F) and 40mBar for 5-10 minutes drying time, because these requirements do not apply in the United States.

Firm Action

On 9/17/2025, correction notices were mailed to customers and distributors who were informed of the following: - The only sterilization requirement now shown on the new Rev D IFU is 4 mins exposure time at 132-Degrees C (270-Degrees F) and 20-30mins drying time. - All affected versions of the IFU should be immediately discarded. - You can access the e-IFU at: https://fusemedical.com/wp-content/uploads/2025/09/Fuse-ULTRA-Foot-Plating-System-IFU-Rev-D-9.2025.pdf - Please pass on this Medical Device Correction to all users of the products and all other people who need to be aware within your organization. - If you have or may have distributed the products listed, please identify and promptly notify those recipients with a copy of this notice, response form, and updated IFU. - Complete and return the response form via email to [email protected] If you have any questions, contact the firm at [email protected]

Root Cause

Labeling design

Distribution Pattern

US Nationwide distribution in the states of NV, TX, LA, WA, WI, FL.

Official Source

View FDA Recall →