Open, Classified

MAC VU360 Acquisition Trunk Cable and Module Holder, high performance, multichannel resting electrocardiograph, Model Number 2030360-001, Product Codes VU2 and VU4

FDA Recall Z-0231-2026 · Posted 2025-10-22

Recall StatusOpen, Classified
Product CodeDPS
Date Initiated2025-08-22
Product Quantity29,225 units

Reason for Recall

If a user incorrectly places the Acquisition Module into the Acquisition Module Holder, it can result in excessive bending and wear of the Acquisition Trunk Cable. This is visible as a damaged sheath or exposed wires. If a damaged Acquisition Trunk Cable is not recognized and removed from service, the cable can overheat with continued use.

Firm Action

An "URGENT MEDICAL DEVICE CORRECTION" notification letter dated 8/22/25 was sent to customers. Actions to be taken by Customer/User: You can continue to use your device in accordance with the instructions in the MAC VU360 Operator Manual and the actions described below: 1. Inspect all potentially affected devices for the presence of damage to the Acquisition Trunk Cable per the MAC VU360 Operator Manual (P/N 2088531-202, Section 11). 2. If the Acquisition Trunk Cable shows damage, remove the cable from service. 3. Ensure the Acquisition Module is placed correctly in the Acquisition Module Holder and that the Acquisition Trunk Cable is not pinched during insertion. See section 1.3 of the MAC VU360 V1.03 Setup and Configuration Manual (P/N 2088531-204). Note: Manuals can be accessed from the GE HealthCare Customer Documentation Portal at https://www.gehealthcare.com/support/manuals 4.Please complete and return the attached acknowledgement form to [email protected] GE HealthCare will provide a customer-installable kit containing a new Acquisition Trunk Cable and a redesigned Acquisition Module Holder for the affected devices at no cost to you. Please make sure all potential users in your facility are made aware of this safety notification and the recommended actions. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution: US (nationwide) and OUS (foreign/Global) to countries of: Australia, Austria, Bahrain, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Korea, Republic of, Kuwait, Malaysia, Netherlands, New Zealand, Norway, Philippines, Puerto Rico, Qatar, Saudi Arabia, Singapore, Sweden, Switzerland, Thailand, United Arab Emirates, and United Kingdom.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.