Oral Probe (Product code 02893-000), accessory used with Welch Allyn SureTemp Plus 690 (Product codes 01690-200 and 01690-400) Thermometers and Welch Allyn SureTemp Plus 692 (01692-200) thermometers
FDA Recall Z-0233-2026 · Posted 2025-10-23
Advertisement
Recalling FirmBaxter Healthcare Corporation
Recall StatusOpen, Classified
Product CodeFLL
Date Initiated2025-09-17
Product Quantity5209
Reason for Recall
Affected oral/axillary probes were inadvertently programmed with the rectal probe configuration. Affected probes will potentially display a lower temperature than the true value, which may lead to a delay in managing fever-related symptoms or infection.
Firm Action
Beginning September 17, 2025, firm notified impacted distributors via Medical Device Recall letters. Customers were instructed to return any affected product and submit a request for replacements. Distributors were instructed to notify their end users via a provided Important Product Information Letter.
For general questions regarding this action, please contact Welch Allyn technical service at 800-535-6663.
Select the following sequence of options: 2 (technical support), 1(English), 1 (physical exam devices), 3 (thermometry), between the hours of 8:00 AM and 8:00 PM Eastern Time, Monday through Friday.
Root Cause
Process control
Distribution Pattern
Distribution US nationwide, Canada and Colombia.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.