VariSoft infusion set, Model Number 1002830
FDA Recall Z-0234-2024 · Posted 2023-11-21
Advertisement
Recalling FirmUnomedical A/S
Recall StatusOpen, Classified
Product CodeFPA
Date Initiated2023-10-11
Product Quantity3362 units
Reason for Recall
The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing.
which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.
Firm Action
Unomedical a/s issued an Urgent Field Safety Notice to its sole consignee on 10/11/2023 via email. The notice explained the issue, potential risk to patient, and requested the distributor to quarantine all affected product pending return and to notify their customers.
Root Cause
Process change control
Distribution Pattern
US Nationwide distribution in the state of CA.