Accutorr 3 Vital Signs Monitor with NIBP valve assembly board, p/n 115-017679-00, containing software version 8.1.5 - Product Usage: The monitor is intended for spot-check monitoring physiologic parameters, including Pulse Oximetry (SpO2), Pulse Rate (PR), Non Invasive Blood Pressure (NIBP), and Temperature (TEMP) on adult, pediatric, and neonatal patients in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians

FDA Recall Z-0235-2020 · Posted None

Recall StatusTerminated
Product CodeMWI
Date Initiated2019-10-01
Date Terminated2021-08-06
Product Quantity13 units

Reason for Recall

NIBP valve assembly board subject to this recall includes enhancements that are not included in the current marketing clearance:1. Anti-motion feature 2. NIBP hose blockage identification function

Firm Action

Mindray communicated by letter on 10/1/19 to consignees and distributors by certified postal mail with return receipt verification: to consignees that have affected vital sign monitor; communication to consignees that purchased directly NIBP valve assembly boards; communication to distributors that have affected vital sign monitor. Consignees were instructed to contact Mindrays Repair Center to arrange for affected units to be returned for a software update.

Root Cause

No Marketing Application

Distribution Pattern

US Nationwide distributions.