3M Attest 1264 Biological Indicators, in boxes of 100 per box
FDA Recall Z-0236-04 · Posted 2003-12-18
Advertisement
Recalling Firm3m Health Care
Recall StatusTerminated
Product CodeFRC
Date Initiated2003-10-24
Date Terminated2005-02-09
Product Quantity9592 boxes of 100 biological indicators per box
Reason for Recall
The 3M Attest 1264/1264P Biological Indicators of ETO sterilization contain a microbiological contaminant which can affect the performance of the positive control and the indicators to some limited extent. The color on a positive control may revert to negative after 24 hours of incubation.
Firm Action
The firm conducted a field correction which began on 10/24/03. They provided revised labeling for the product. Letters, dated October 24, 2003, were sent to consignees. The letters had a description of the problem, instructions for doing the relabeling in the field, and recommendations for users of the product.
Root Cause
Other
Distribution Pattern
The product was distributed nationwide in the United States and worldwide.