Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

OMNI K2 Hip Stem, Manufacturer's Product Code 130006 The OMNI K2 Hip Stem is intended for use as the femoral component of a primary total hip replacement. This femoral hip stem is intended for uncemented fixation and single use implantation.

FDA Recall Z-0237-2022 · Posted None

Recalling FirmOmniLife Science
Recall StatusTerminated
Product CodeLPH
Date Initiated2021-10-01
Date Terminated2022-07-20
Product Quantity7

Reason for Recall

The device is missing grit blasting on the proximal surface of the femoral stem, which may lead to the possibility of product revision in the event of undersizing.

Firm Action

Affected agents and representatives were notified by email, with initial communication sent via a quarantine form on 10/1/21. Additional notices titled "Urgent Medical Device Recall" were emailed out on October 15, 2021. Customers were instructed to immediately quarantine lot 38933 and return the product to OMNI. Customers were provided with a response form and asked to complete and return the form to the recalling company by FAX to 508-819-3390 or my email at [email protected]. For further information contact Christina Rovaldi, RA/QA Manager at 774-226-1847 or 800-448-6664. Company contact hours are Monday through Friday, 8:00 AM to 4:30 PM, Eastern Time.

Root Cause

Process control

Distribution Pattern

US Distribution to states of: Colorado and Virginia.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑