Terminated

UniCel DxH Slidemaker Stainer, Part Number 75222, for use with automated hematology analyzer

FDA Recall Z-0238-2018 · Posted None

Recalling FirmBeckman Coulter Inc.
Recall StatusTerminated
Product CodeGKZ
Date Initiated2017-11-03
Date Terminated2018-06-19
Product Quantity3 units

Reason for Recall

The shield for the Single-Tube Station on the DxH instrument may interfere with sample aspiration in the open vial position.

Firm Action

The firm notified its consignees of the recall on 11/03/2017 by letter. The letter stated the issue and provided the following instructions: "To avoid a biohazardous spill, do not use the right (green) position of the Single-Tube Station for sample aspiration until the shield is removed. You may remove the shield following the instructions in Attachment 1 or contact your local Beckman Coulter Representative for further assistance."

Root Cause

Device Design

Distribution Pattern

CA, AR, HI, CA, OH, WA, MD, Canada, Japan