Philips Azurion systems with software releases 2.1(L1) and 2.1(L2)

FDA Recall Z-0238-2022 · Posted None

Recalling FirmPhilips Healthcare
Recall StatusOpen, Classified
Product CodeOWB
Date Initiated2021-09-24
Product Quantity226 units

Reason for Recall

3D-RA is a reconstruction software product that can be used with the Philips Azurin system. This product extends the functionality of compatible X-ray equipment with the capability of 3D X-ray imaging. Before starting a 3D-RA scan, the detector of the Philips Azurion system must be positioned in portrait or landscape orientation with an angle of 0 or +/-90 degrees relative to the 3D scan direction. Philips has discovered that Azurion releases 2.1(L1) and 2.1(L2) allow creation of a 3D-RA scan with a deviating detector orientation of up to 3 degrees from the exact portrait or landscape orientation (i.e., at 0 or +/-90 degrees), relative to the 3D scan direction. 3D-RA scans with a deviating detector orientation of ore than 1.0 degree from the exact portrait or landscape orientation (i.e., at 0 or +/-90 degrees) cannot be reconstructed with the 3D-RA reconstruction software.

Firm Action

Philips will inform customers of the issue via an Electronic Product Radiation Correction notification. The letter identified the affected product, problem and actions to be taken. Philips Field Service Engineers will schedule an appointment with customers to install the software update. For questions contact your Philips representative at 1-800-722-9377.

Root Cause

Radiation Control for Health and Safety Act

Distribution Pattern

US Nationwide distribution