Fogarty Thru-Lumen Embolectomy Catheter, REF: 12TLW805F35

FDA Recall Z-0238-2024 · Posted 2023-11-06

Recall StatusOpen, Classified
Product CodeDXE
Date Initiated2023-09-07
Product Quantity185

Reason for Recall

Inadvertent release of embolectomy catheters that may experience slower than normal balloon deflation (greater-than 15 seconds).

Firm Action

On 9/7/23, correction notices were sent to customers informing them to take the following actions: 2) Return affected devices. 3) Forward notice to any of your customers who have purchased the impacted devices. 3) Complete and return the acknowledgement form.

Root Cause

Nonconforming Material/Component

Distribution Pattern

OUS: Singapore, Taiwan, Canada