Computed Tomography Systems Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition
FDA Recall Z-0242-2022 · Posted 2021-11-24
Advertisement
Recalling FirmGE Healthcare, LLC
Recall StatusOpen, Classified
Product CodeJAK
Date Initiated2021-06-30
Product Quantity105 in total
Reason for Recall
The accumulated dose is incorrectly displayed in certain situations. Thisoccurs during an interventional exam using the SmartStep option if the table height is adjusted after
the exam starts. In this situation, the SmartStep display will show a value that is higher (up to 100
times) than the actual dose delivered.
Firm Action
GE sent an Important Electronic Product Radiation Warning Notification letter dated October 4, 2021 to customers, The letter identified the affected product, problem and actions to be taken. Customers were informed that GE Healthcare will, without charge, remedy the defect or bring the product into
compliance with each applicable Federal standard in accordance with a plan to be
approved by the Secretary of Health and Human Services, the details of which will be included in a subsequent communication to you or through a GE Healthcare field engineer site visit. For questions contact GE
Healthcare Service at 1-800-437-1171 or your local Service Representative.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
Worldwide and US Nationwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.