Sedecal Mobile Digital Diagnostic X-Ray Systems; SM-40HF-B-D-KM (AeroDR X30)

FDA Recall Z-0242-2026 · Posted 2025-10-23

Recalling FirmSEDECAL SA
Recall StatusOpen, Classified
Product CodeIZL
Date Initiated2025-06-10
Product Quantity56 units

Reason for Recall

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

Firm Action

Sedecal emailed their distributors on 06/10/2025 to remind them the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning. on 09/19/2025 the firm issued an Important Safety Notice to its direct consignees via email. This notice initiates Phase 2 of the recall strategy expanding the scope and providing information on stickers to cover gaps around the fine positioning control buttons and a DMC software update that stops unit movement if triggered by a short cicuit or prolonged button press with instructions. Consignees will be contacted by a Sedecal Customer support Service to coordinate the shipment of the necessary conversion kits (quantity and delivery address) to allow upgrading the affected units.

Root Cause

Device Design

Distribution Pattern

US: CA, IL, and NJ