Ilizarov Wire Tensioner, Model 71070341, UDI (10)08033201842072(10)21HP02550(11)210831(21E434
FDA Recall Z-0245-2022 · Posted None
Advertisement
Recalling FirmMedicalplastic S.R.L.
Recall StatusOpen, Classified
Product CodeKTT
Date Initiated2021-06-17
Product Quantity581 units
Reason for Recall
Medicalplastic s.r.l. is conducting a voluntary field action to remove a group of Ilizarov Wire Tensioners due to a manufacturing error. An internal washer of the Ilizarov Wire Tensioner has been made in a wrong plastic material that could wear and break during the sterilization process in autoclave. Consequently this could cause malfunctioning of the tensioner.
Firm Action
The firm notified its sole consignee of the issue and its intent to recall by email on 06/17/2021. The recalled washers will be replaced.
Root Cause
Process control
Distribution Pattern
Memphis, TN