Terminated

TherMax Blood Warmer Unit

FDA Recall Z-0247-2020 · Posted 2019-10-30

Recall StatusTerminated
Product CodeKDI
Date Initiated2019-09-19
Date Terminated2020-12-11
Product Quantity504 units total

Reason for Recall

TherMax Blood Warmers may not be in compliance with an electrical safety standard, which requires a protective earth connection that can sustain a current of 25 amps for 10 seconds.

Firm Action

Baxter Healthcare notified customers on about 09/19/2019, via "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were informed that specific TherMax Blood Warmers listed below may not be in compliance with an electrical safety standard, which requires a protective earth connection that can sustain a current of 25 amps for 10 seconds. The affected units may not have an adequate ground path, which could cause the units to become damaged if used with improperly grounded outlets. The damage could cause the TherMax Blood Warmer units to be nonfunctional during clinical use. Instructions included that customers could continue to safely use affected devices with a properly grounded outlet, that a local Baxter representative will contact each facility to arrange for the correction of affected devices, to complete and return the Baxter Customer Reply Form via fax to 224-270-5457 or scanning and e-mailing it to [email protected], and to notify customers if the product was further distributed. For general questions regarding this communication, contact Baxter Corporate Product Surveillance at 800-437-5176, between the hours of 8:00 am and 5:00 pm Central Time, Monday through Friday.

Root Cause

Device Design

Distribution Pattern

Nationwide distribution to Georgia and Maryland. International distribution to Belgium, Germany, Finland, France, Greece, Italy, Luxemburg, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Kingdom, Australia, and New Zealand