Level 1 Normothermic I.V. Fluid Administration Set REF DI-60HL (Foreign Distribution Only)

FDA Recall Z-0251-2007 · Posted 2006-12-06

Recall StatusTerminated
Product CodeFPA
Date Initiated2006-11-07
Date Terminated2009-11-30
Product Quantity13,640 units

Reason for Recall

IV Fluid Administration Set may have a blockage of the recirculating warming fluid channel within the disposable administration set, causing an alarm condition of over-temperature during priming.

Firm Action

Smiths Medical ASD notified Direct customers and distributors via letter sent by UPS on or about November 6, 2006 and the days following as necessary until all consignees are notified. Letters to direct customers will be addressed to the Risk Manager. All Smiths Medical affiliates contacted by e-mail and given instructions to send a letter to their customers and to confirm to Smiths Medical that their customers have taken appropriate action by returning the affected product. US Distributors instructed to return the affected product and to provide a customer list. Foreign distributors instructed to return the affected product and given instructions to send a letter to their customers and to confirm to Smiths Medical that their customers have taken appropriate action by returning the affected product.

Root Cause

Equipment maintenance

Distribution Pattern

Worldwide, including USA, Japan, Canada, Ireland, UK, Finland, Netherlands, Germany, Austria, Portugal, France, Sweden, Australia, New Zealand, Russia, Italy, Singapore, Kuwait, Saudi Arabia, Czech Republic, Dubai, Spain, Norway, Denmark, and East Temor.