BD Vacutainer Luer-Lok Access Device Holder with Pre-Attached Multiple Sample Adapter Sterile, Do Not Reuse, Keep away from heat, Rx only Ref # 364902 200 (4X50) BD Franklin Lakes, NJ 07417
FDA Recall Z-0253-2007 · Posted 2006-12-06
Advertisement
Recalling FirmBecton Dickinson & Company
Recall StatusTerminated
Product CodeJKA
Date Initiated2006-09-08
Date Terminated2007-05-08
Product Quantity1,744,400
Reason for Recall
Reports of failure of the Luer Lok Access Device to lock securely to certain catheter devices because the male luer taper surface is not within the specification.
Firm Action
Recall notification was sent 9/8/06 to all direct accounts and distributors by US mail, return receipt requested. Consignees were asked to send product back to BD with the packing slip and shipping labels enclosed
Root Cause
Other
Distribution Pattern
Worldwide, including USA, Canada, and Australia.