Amplatzer Cardiac Septal Occluder Delivery Systems labeled as follows: a) AMPLATZER Delivery System, Order no. 9-DEL-5F-180/60, 9-DEL-6F-45/60, 9-DEL-6F-180/80, 9-DEL-7F-45/60, 9-DEL-7F-45/80, 9-DEL-7F-180/80, 9-DEL-8F-45/60, 9-DEL-8F-45/80, 9-DEL-8F-180/80, 9-DEL-9F-45/80, 9-DEL-9F-180/80, 9-DEL-10F-45/80, 9-DEL-12F-45/80. [Breakdown of order no. ''9-DEL-5F-180/60'' is as follows: 9-DEL, French size (5F) - degree angle curve (180) / usable length in cm (60).] b) the following devices are not approved for US distribution: AMPLATZER TorqVue Delivery System, order no. 9-ITV06F45/60, 9-ITV07F45/60, 9-ITV07F45/80, 9-ITV08F45/60, 9-ITV08F45/80, 9-ITV09F45/80, 9-ITV10F45/80, 9-ITV12F45/80, 9-ITV05F180/60, 9-ITV06F180/80, 9-ITV07F180/80, 9-ITV08F180/80, 9-ITV09F180/80. Manufactured in USA, Sterile EO. Headquarters AGA Medical Corporation, 682 Mendelssohn Avenue, Golden Valley, MN 55427 U.S.A. The Delivery System is comprised of a delivery sheath, delivery cable, dilator, loading device and pin vise.

FDA Recall Z-0255-2007 · Posted 2006-12-19

Recall StatusTerminated
Product CodeGBK
Date Initiated2006-10-02
Date Terminated2008-10-09
Product Quantity90,516

Reason for Recall

AGA Medical has found that microscopic tears can occur in the delivery system sterile packaging under accelerated stress testing with routine shipping configurations. These microscopic tears are a potential breach of the sterile barrier. AGA has no confirmed complaints or adverse events related to this failure mode.

Firm Action

An Urgent Recall Notice, dated 09/29/06, was sent to customers via facsimile or email and via Federal Express. The letter describes the issue and what product is affected. Unexpired product that is returned to AGA will be replaced. A response form is asked to be returned to AGA.

Root Cause

Other

Distribution Pattern

Worldwide, including USA, Puerto Rico, Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Cuba, Denmark, El Salvador, Finland, France, Germany, Greece, Guatemala, Hungary, Iceland, India, Indonesia, Israel, Italy, Japan, Jordan, Korea, Kuwait, Malaysia, Moroc, Mexico, Norway, Pakistan, Peru, Phillipines, Poland, Portugal, Saudi Arabia, Singapore, Slovak Republic, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, The Netherlands, Tunisia, Turkey, United Kingdom, Uruguay, Venezuela, and Vietnam.