Terminated

BioPlex 2200 System, Cat. No. 665-1460A containing APF CD SW2_v1., Syphilis IgG IFU Packet contains the Syphilis IgG Assay Protocol File LAPF CD

FDA Recall Z-0256-2008 · Posted 2007-11-09

Recall StatusTerminated
Product CodeLIP
Date Initiated2007-09-25
Date Terminated2008-04-16
Product Quantity36

Reason for Recall

Mislabeling: Reporting terminology in the Syphilis IgG APF CD is not consistent with the distributed Instructions for Use. (Non-Reactive and Reactive rather than Negative and Positive)

Firm Action

The recalling firm notified consignees by phone call on 9/25/2007, followed up with a fax notification. The firm plans to monitor the consignees that do not respond by tracking on a spreadsheet and following up with another phone call. Consignees are instructed to destroy the APF CD which will be replaced at a later date.

Root Cause

Labeling Change Control

Distribution Pattern

Nationwide

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →