Sysmex XE-2100C Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe, Japan; Sysmex America, One Nelson C. White Parkway, Mundelein, IL 60060; catalog #053-2311-4 The Sysmex XE-2100C is an automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories.
FDA Recall Z-0256-2013 · Posted 2012-11-06
Advertisement
Recalling FirmSysmex America, Inc.
Recall StatusTerminated
Product CodeGKZ
Date Initiated2012-06-01
Date Terminated2016-08-22
Product Quantity69 units
Reason for Recall
There is a variation of reticulocyte counts between XE Series (XE-2100, XE-5000) and XT-Series (XT-2000i, XT-4000i) hematology analyzers that may result in a retic counts that are an average of 21% lower on the XE-Series because of a low bias.
Firm Action
Sysmex America, Inc. sent a "PRODUCT NOTIFICATION" letter dated May 2012 to all affected customers on June 1, 2012. The letter identifies the product, problem, and actions to be taken. The letter instructs customers to review patient results to determine if their Reference Intervals need adjustment, and to review QC results via Insight and document any change that may be linked to the date of calibration if appropriate.
Any questions concerning the information contained in the notification were directed to the Sysmex Technical Assistance Center at 1-888-879-7639.
Root Cause
Device Design
Distribution Pattern
Nationwide distribution: USA including Puerto Rico.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source