Veris Invasive Blood Pressure Interface Cables-Part # 3010478. The product is an accessory to patient physiological monitor.

FDA Recall Z-0259-2007 · Posted 2006-12-15

Recalling FirmMedrad Inc
Recall StatusTerminated
Product CodeMWI
Date Initiated2006-09-29
Date Terminated2007-06-20
Product Quantity80 units

Reason for Recall

Interference problems: certain invasive blood pressure transducers connected to these cables are susceptible to interference if exposed to static magnetic field strengths exceeding 300 gauss. In certain orientations, the transducer causes the invasive blood pressure readings on the physiological monitor to display inaccurate readings.

Firm Action

The recalling firm contacted all affected customers via telephone or letter dated November 20, 2006 to inform them of the problem and instruct them to return the cables or apply corrective labeling.

Root Cause

Other

Distribution Pattern

Worldwide distribution ---- including USA state of CA, CO, HI, IL, IN, MA, MD, NJ, NY, OH, PA, TX, and WA and countries of The Netherlands, Denmark, Australia, Brazil, Italy, and Japan.