Custom 5.5 mm Bivona Uncuffed Hyperflex Tracheostomy Tube, I.D. 5.5 mm, O.D. 8.0 mm, length 4.75 mm, Mfrd by Smiths Medical Critical Care; REF HU05JS55NNA112S.

FDA Recall Z-0260-2007 · Posted 2006-12-07

Recall StatusTerminated
Product CodeJOH
Date Initiated2006-11-17
Date Terminated2007-01-12
Product Quantity5

Reason for Recall

Mislabeled as to size: The product is 47.5 mm in length, but the label lists the length as 4.75 mm.

Firm Action

The consignee was notified of the problem via recall letter sent on 11/17/06 and given the options of continuing to use the product or of returning it.

Root Cause

Other

Distribution Pattern

New York-single hospital consignee