Stryker Orthopaedics Tapers LFIT V40 Vitallium femoral heads. hip prosthesis component.
FDA Recall Z-0261-2016 · Posted 2015-11-10
Advertisement
Recalling FirmStryker Howmedica Osteonics Corp.
Recall StatusTerminated
Product CodeLPH
Date Initiated2015-09-08
Date Terminated2017-07-25
Product Quantity201
Reason for Recall
Complaints that the femoral head could not be assembled with its corresponding V40 stem trunnion at the time of surgery.
Firm Action
Stryker sent an Urgent Medical Device Recall Notifiications dated September 9 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
Customers are asked to complete the attached form and return within 5 days by using email [email protected] or FAX 1-888-965-5803.
For further questions please call ( 201) 831-5000.
Root Cause
Process design
Distribution Pattern
Nationwide Distribution