Terminated

The ARCHITECT i2000, Clinical Chemistry Analyzer, Manufactured by Abbott Laboratories, Irving, Texas

FDA Recall Z-0262-04 · Posted 2003-12-24

Recall StatusTerminated
Product CodeJQP
Date Initiated2003-11-17
Date Terminated2004-10-28
Product Quantity223

Reason for Recall

Sample Identification (SID) numbers of a bar coded sample may be truncated, or shortened, after being scanned by the bar code reader (e.g. an actual SID of 123456 is shortened to 1234)

Firm Action

The firm plans to distribute a Product Information Letter and an updated ARCHITECT i System Operations Manufal Addendum to all US ARCHITECT customers. An electronic version of these documents was sent to all Abbott affilities worldwide for distribution to their ARCHITECT customers.

Root Cause

Other

Distribution Pattern

Mexico, Puerto Rico, Japan, Taiwan, Australia, Colombia, Cananda, Hong Kong, New Zealand, Netherlands, Singapore, Germany, Grand Cayman, Korea, Chile, Thailand, TX, PA, SC, CA, IN.