Rezum System, Vapor BPH Ablation System, Model Number D2201-003. NxThera Inc. 7351 Kirkwood Lane Suite 138 Maple Grove, MN 55369 USA. Intended to relieve symptoms, obstructions, and reduce prostate tissue associated with benign prostatic hyperplasia (BPH). It is indicated for men greater or equal to 50 years of age with a prostate volume greater than or equal to 30cm3 and less than or equal to 80cm3. The Rezum System is also indicated for treatment of prostates with hyperplasia of the central zone and/or a median lobe.

FDA Recall Z-0262-2018 · Posted None

Recalling FirmNXTHERA
Recall StatusTerminated
Product CodeKNS
Date Initiated2017-07-28
Date Terminated2018-08-01
Product Quantity113 units total

Reason for Recall

Needle Bond may not be able to withstand the tension force applied when the shaped needle is deployed into and retracted from prostate tissue to deliver treatments.

Firm Action

Customers were notified initially by sales representatives on or about July 18, 2017. Any affected devices were to be returned to NxThera. Additionally, a formal recall letter was mailed to customers on approximately July 28, 2017. NxThera's telephone contact information and FEDEX number will be provided on the verification form. For further questions, please call (763) 515-2084..

Root Cause

Nonconforming Material/Component

Distribution Pattern

USA Distribution to the states of : AL, AZ, CA, CO, FL, GA, IL, MA, MI, MN, MO, NE, NJ, NY, OH, PA, TN, UT, WA.