SYNCHRON Clinical Systems Lipase (LIPA) Reagent; Classification name: 862.1465
FDA Recall Z-0262-5 · Posted 2004-11-30
Advertisement
Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeCHI
Date Initiated2004-10-25
Date Terminated2005-06-17
Product Quantity6,698
Reason for Recall
Beckman Coulter has confirmed that occasionally a cuvette may be skipped during the SYNCHRON Lipase Wash (LIWA) procedure. If this occurs, there is the potential for carryover into the SYNCHRON Lipase (LIPA) assay, resulting ina substantial positive bias affecting Lipase (LIPA) results.
Firm Action
The week of 10/25/2004, a product corrective action (PCA) letter was mailed vis U.S. mail to all customers shipped LIPA reagent for the 24 months preceding the discovery of the issue. Customers instructed to repeat any sample that produces a lipase value of > 170 U/L that cannot be confirmed by patient history.
Root Cause
Other
Distribution Pattern
US and Canada
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.