Terminated

Stryker Instruments 2.1 mm Cross-Cut Fissure Bur, carbide extended, sterile; REF 2296-101-621.

FDA Recall Z-0263-2010 · Posted 2009-11-16

Recall StatusTerminated
Product CodeDZI
Date Initiated2009-10-05
Date Terminated2009-11-30
Product Quantity93

Reason for Recall

The cross-cuts are missing on the device, which will result in poor cutting and overheating and/or smoking of the device.

Firm Action

Consignees were notified by letter dated 10/5/09 and requested to return the devices to Stryker. The international account was notified by email.

Root Cause

Vendor change control

Distribution Pattern

USA (California, Georgia, Louisiana, Ohio, Wisconsin) and Brazil.