Infinity(R) Acute Care System (IACS) Monitoring Solution (M540) - Physiological Patient Monitoring Solution Networked solution/system used to monitor patients' vital signs and therapy, control alarms, review Web-based diagnostic images, access patient records, and scale up functionality based on patient acuity level.
FDA Recall Z-0263-2012 · Posted 2011-11-22
Advertisement
Recalling FirmDraeger Medical, Inc.
Recall StatusTerminated
Product CodeMHX
Date Initiated2011-10-18
Date Terminated2013-04-01
Product Quantity476
Reason for Recall
During product training on the Infinity(R) Acute Care System (IACS) Monitoring Solution (M540) it was discovered that the weight-based drug dosage calculation may indicate incorrect recommended values. In addition, a misalignment of ECG and ART waveforms delivered from an Infinity(RO Acute Care System (IACS) Monitoring Solution (M540) was observed on the Infinity Central Station.
Firm Action
The firm initiated their recall of this device on October 17, 2011 by issuing a letter to the consignee instructing customers to follow the Instructions for Use and use the Infinity Central Station only for remote assessment of a patient's status. They recommend that you do not consider the waveforms on Infinity Central Station for primary diagnosis purposes. In case of a high priority alarm situation, "Check the patient and treat if necessary".
If the misalignment issue is observed, it can be cleared by undocking and redocking the M540 or by changing the order of waveforms displayed on the Cockpit screen (or any other change to the auto display screen layout).
Root Cause
Software design
Distribution Pattern
Nationwide distribution - IL
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.