Autoject EI, REF AJ1310
FDA Recall Z-0263-2020 · Posted None
Advertisement
Recalling FirmOwen Mumford USA, Inc.
Recall StatusOpen, Classified
Product CodeKZH
Date Initiated2019-07-17
Product Quantity1510 units
Reason for Recall
There is a possible assembly error on Autoject EI, lot number V14. This lot was packed into two products: AJ 1310, lot number V14 and AN 3310VET (branded as Petfine Auto Injector), batches V14.V976 & V14.W107
Firm Action
The firm initiated the recall by letter on 07/17/2019. The consignee was directed to quarantine all affected product and contact the company for return. Distributors were also directed to cease distribution and contact their customers.
Root Cause
Employee error
Distribution Pattern
FL, LA, SC, CA, NJ, TX, NC, CO, NH, KY, NY, MO, PA