Terminated

Stryker Endoscopy 3.55 MM Super 90 S-SERFAS Energy Probe; electrosurgical device. Model number: 279-351-300;

FDA Recall Z-0265-2012 · Posted 2011-11-21

Recalling FirmStryker Endoscopy
Recall StatusTerminated
Product CodeGEI
Date Initiated2011-10-24
Date Terminated2012-02-29
Product Quantity6307

Reason for Recall

An increased number of complaints for tip breaks were observed for the 3.5mm Super 90-S SERFAS Energy Suction Probes manufactured from June to August 2011.

Firm Action

Stryker Endoscopy Urgent Medical Device Removal notifications datyed October 25, 2011, were sent to all consignees via traceable mail. Sales representatives and regional managers were notified by e-mail. Product is to be returned to Stryker Endoscopy using the pre-paid shipping label. Replacement product may be obtained by calling Stryker at 1-800-624-4422. The enclosed Acknowledgement receipt form is to be completed and returned. Questions or comments may be directed to Stryker at 1-800-624-4422 or e-mail at [email protected].

Root Cause

Component design/selection

Distribution Pattern

Worldwide distribution including USA, Australia, Canada, China, Greece, Hong Kong, India, Italy, Japan, France, Netherlands, Poland, South Africa, Spain, Sweden, Switzerland and the United Kingdom

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.