MODULARIS VARIOSTAR; Lithotripter device designed to treat urolithiasis.
FDA Recall Z-0265-2016 · Posted None
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeLNS
Date Initiated2015-09-24
Date Terminated2016-12-05
Product Quantity15
Reason for Recall
Display freeze of MODULARIS hand control results in information not being updated on the display. Current treatment data is not shown to the user.
Firm Action
Siemens sent a Field Safety Notice (XP044/15/S) dated September 24, 2015 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.
Root Cause
Software design
Distribution Pattern
Distributed in PR and the states of MO, NC, GA, MS, LA, and KY.