Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Nasal Mucosal Atomizer Device with syringe, LATEX FREE For Single Patient Use Only Packaged Clean -Non-Sterile- MADE IN CANADA. REF: LTAD-N1-25, Lot 121608, packed 25 per package. REF: LTAD-N1-100, Lot 121608, packed 100 per package. Manufactured By: Southmedic 50 alliance Blvd, Barrie, Ontario, Canada L4M 5K3. Distributed By: Sharn Inc. Anesthesia 1-800-325-3671 813-886-2701 email: [email protected] The device is used to deliver a fine mist numbing agent to the nose or throat for intubating a patient immediately prior to undergoing anesthesia.

FDA Recall Z-0266-2010 · Posted 2009-11-16

Recalling FirmSharn, Inc.
Recall StatusTerminated
Product CodeCCT
Date Initiated2009-03-05
Date Terminated2012-07-13
Product Quantity550

Reason for Recall

Sharn Anesthesia Inc., Tampa, FL received reports of leaking containers following use of their Sharn Anesthesia Microblast nasal and pediatric laryngeal atomizer device.

Firm Action

Recalling firm is contacting all direct accounts (distributors & user facilities) by telephone and letter. They request completion and return of the of the Return Goods Form and of any affected product by UPS using the firm's account number.

Root Cause

Device Design

Distribution Pattern

Nationwide Distribution

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑