The GMK knee system: designed for cemented use in total knee arthroplasty. Sphere and Primary Inserts with part number ranges: 02.12.0110FL - 02.12.0620FL; 02.12.0110FR - 02.12.0620FR; 02.07.0110PSF - 02.07.0620PSF

FDA Recall Z-0266-2018 · Posted 2017-12-20

Recalling FirmMedacta Usa Inc
Recall StatusTerminated
Product CodeJWH
Date Initiated2017-10-24
Date Terminated2018-06-20
Product Quantity12,768 units

Reason for Recall

Medacta has observed a worldwide occurrence rate of post-operative insert screw back-out of approximately 0.1%. For this reason, we stress the importance of using the 3.5 Nm torque limiting driver if the screw is used, as directed in our surgical techniques.

Firm Action

All surgeon customers shall be notified via email by October 24, 2017 of the reminder to follow the surgical technique. To ensure that surgeons receive the notification, we will send the letter via email so that we can track and ensure the message was received. If the email is returned, we would then send a personalized letter directly to the surgeon. A copy of this notification is included with the submission of this report.

Root Cause

Device Design

Distribution Pattern

nationwide