Wolf Medical Supply, Inc., WOLF-PAK REDI-FLO ELASTOMERIC INFUSION PUMP, FILL VOLUME 60 mL, FLOW RATE 2mL/Hr, Model # CT-0020-060 Item # RFC060020. Manufactured by Medpro International (Thailand) Ltd.

FDA Recall Z-0267-2012 · Posted 2011-11-23

Recall StatusTerminated
Product CodeMEB
Date Initiated2011-10-20
Date Terminated2012-04-25
Product Quantity8 cases (192 units)

Reason for Recall

Redi-Flo Elastomeric Infusion Pumps may have a higher flow rate than specified (more than +15%).

Firm Action

On 10/20/2011 Wolf Medical Supply Inc., customers were notified by telephone and email. Product was distributed to end users.

Root Cause

Process control

Distribution Pattern

Nationwide distribution, including Puerto Rico.