Patient tables of CT systems with brand name: CT/e; CT/e Dual; CT/e Plus, CT/e Dual Plus; CT/e Lite; ProSpeed AI; ProSpeed AII; ProSpeed FI; ProSpeed FII; ProSpeed EII; CT HiSpeed Series.

FDA Recall Z-0269-05 · Posted 2005-02-24

Recall StatusTerminated
Product CodeIXR
Date Initiated2004-10-21
Date Terminated2006-07-24
Product Quantity1150

Reason for Recall

The patient tables of the affected CT systems are supported by a hydraulic cylinder and a cylinder shaft (i.e. the upper shaft) which connects the tabletop to the table frame. GE received a report of a HiSpeed LX/I system table falling abruptly due to the filure of the upper shaft. No injuries were reported.

Firm Action

Inadequate materials used to manufacture the upper shaft of patient tables that can result in uncontrolled table motion.

Root Cause

Other